Diffusion Methanometers
In the latest types of methanometers the air being sampled reached the sensing element by diffusion. Therefore it is only when drawing a sample through a probe that an aspirator is used.
THE M.S.A. D6 METHANOMETER
To operate the D6, push up the yellow dust cover to expose the diffusion head. Press the methane test button (the larger of the two buttons). This allows current to flow through the pellistor; its amount depending on the concentration of methane. Keep the button depressed until a steady reading is obtained. Then replace the dust cover.
To check the battery voltage press both buttons. After five seconds release the methane test button. If the battery is charged, the needle will settle in the red sector between 2.8 and 2.1 on the scale whether or not methane is present. (So long as the battery voltage exceeds 2.3, only the voltage button need be pressed).
Before going underground, a check should be made in fresh air of the battery voltage (as above), and of the zero. For the latter, check that the needle settles on the zero both (a) with neither button pressed and (b) with methane button pressed.
For remote sampling, the correct M.S.A. probe (or M.S.A. remote sampling adaptor with extension tube) must be used. After pushing up the yellow dust cover, screw on the probe till the knurled nut at the base is tight. Slacken the knurled nut at the top, extend the probe, then tighten the nut. Pressing of the methane test button must commence before aspiration commences, in order to distinguish concentrations under 5% from those over 40%. Aspiration should continue till a steady reading is obtained. On completion replace the dust cover.