INSTRUCTIONS FOR USING THE GARFORTH LAMP

1. General

The lamp may be used for testing for firedamp in the same way as any other flame lamp when the area to be tested is in the general body of the air. Where the area to be tested is near the roof or is such that the lamp cannot be positioned so that the flame can be properly observed, or where the firedamp content is likely to exceed 4%, use is made of the aspirator bulb to take a remote sample which can thereafter be conveniently introduced into the lamp. The principle of introducing a sample around the wick allows firedamp concentrations up to about 20% to be estimated by means of gas caps on or around the testing flame.

For practical purposes it is the gas caps indicating concentrations of up to 5% that need to be carefully memorised, since 5% is the lower explosive limit.

2. Sampling with the Aspirator Bulb

The aspirator bulb and valve assembly allow remote sampling to be carried out as follows:

 (a) 
The bulb is removed from the valve assembly, squeezed tightly, folded over and squeezed again in order to expel as much air as possible;
 (b) 
where the sampling position is within reach of the operator, locate the nozzle of the deflated bulb by hand at the sampling point and release the pressure of the hand. Retain the bulb at the sampling point until it has regained its normal shape when it can be removed for introduction into the lamp;
 (c) 
where the sampling position cannot be reached by hand fit the deflated bulb to the valve assembly. After fitting, raise the valve cap momentarily, squeeze the bulb firmly and release the valve cap. The bulb will stay deflated when the pressure of the hand is released. Attach the valve assembly to a stick which allows the bulb to be located at the sampling position. Open the valve by pushing the button against the roof or other solid object or, where no such object exists, by means of wire or string fitted through the hole in the valve cap arm which can be pulled when the bulb is in position. In either case the valve must be held in the open position until the bulb is fully inflated. Remove the bulb from the valve assembly, taking care it is not squeezed in the process. The exact sampling location will be determined by the position of the valve inlet.

3. Introducing the Sample into the Lamp

 (a) 
Adjust the wick height to give a testing flame with the minimum of yellow present;
 (b) 
insert the nozzle of the sampling bulb carefully into the socket in the base or side of the lamp;
 (c) 
raise the lamp to eye level or to the preferred observation position (as the gas appears as soon as the bulb is squeezed the operator should be ready to observe the cap immediately);
 (d) 
squeeze the bulb steadily so that it empties at an even rate in five to seven seconds.
  
 Notes (i) 
If the bulb is emptied in a shorter time the cap shape is not appreciably altered but the viewing time is reduced;
 (ii) 
If the bulb is squeezed too slowly the gas cap is not properly formed and any firedamp present would be underestimated.
 (e) 
although there is little danger of loss of sample from the bulb it is advisable to introduce the sample into the lamp as soon as possible after it is taken.

4. Estimation of Firedamp Concentration

 (a) 
If the sample contains a methane concentration which is not detectable by means of a flame lamp, the testing flame is flattened and the yellow eye is greatly reduced or disappears completely;
 (b) 
if the injected sample contains between 1% and 4½% of firedamp, the gas cap obtained will be similar in shape but slightly smaller than when the lamp is used for detecting firedamp in the general body of the air at the same percentage;
 (c) 
when the injected sample contains more than 4½% firedamp, i.e. at concentrations which cause the normal flame lamp to be extinguished, a flame indication is still obtained. For any concentration up to about 20% firedamp there is a characteristic shape, size and colouring of the flame, and, with some experience estimates of the approximate percentage in the sample can be made;
 (d) 
above 20% firedamp a tall column of flame reaching to above the top of the glass is obtained within a second of starting to squeeze the bulb;
 (e) 
the presence of up to 2% of firedamp in the general body of the air where the lamp is being used does not have any appreciable effect on the gas cap produced by an injected sample.